Meridian Bioscience, Inc.: Magellan Diagnostics, Inc. Expands Voluntary Recall of LeadCare Test Kits
This is a paid press release. Contact the press release distributor directly with any inquiries.

Meridian Bioscience, Inc.: Magellan Diagnostics, Inc. Expands Voluntary Recall of LeadCare Test Kits

CINCINNATI, OH / ACCESSWIRE / September 1, 2021 / Meridian Bioscience, Inc. (NASDAQ:VIVO), a provider of diagnostic testing solutions and life science raw materials, today announced its wholly owned subsidiary, Magellan Diagnostics, Inc. ("Magellan"), expanded the Class I recall of its LeadCare® II Blood Lead Test Kits, LeadCare Plus® Blood Lead Test Kits, and LeadCare Ultra® Blood Lead Test Kits (the "LeadCare Test Kits") for the detection of lead in whole blood.

Magellan provides two controls in the test kits which are designed to mimic blood and are spiked with lead to specific target values with an associated acceptable range. Results of the control tests within the acceptable range indicate that the system is operating properly before testing patient samples. In May 2021, Magellan initiated this voluntary recall after identifying an ongoing issue with testing of the controls included in the LeadCare Test Kits. Magellan continues to investigate this issue and has conducted extensive testing to evaluate potential root causes.

Scope of Recall
Magellan received reports that control tests of either the "Low-Control" (e.g., the "Level 1" control at approximately 9 mg/dL ± 3 mg/dL) and/or the "High-Control" (e.g., the "Level 2" control at approximately 28 mg/dL ± 4 mg/dL) generated a "low" result (i.e., "Control Out of Range-Low" ["COOR-L"]). Magellan initiated the recall because the impacted LeadCare Test Kits lots could potentially underestimate blood lead levels when processing patient blood samples.

As part of the recall to the user level, Magellan is notifying customers and distributors affected by the recall. Magellan's customer recall notification provides instructions for the return and replacement of the impacted LeadCare Test Kits (see list of affected lot numbers below).

Catalog No.

Product

Lot Number

Expiration Date

70-6762

LeadCare® II Blood Lead Test Kit

2012M sub-lots***

Apr 8, 2022

2013M*

Apr 22, 2022

2014M*

Apr 29, 2022

2015M*

May 12, 2022

2016M*

May 19, 2022

2017M*

Jun 10, 2022

2018M

Jun 6, 2022

2101M**

Jul 28, 2022

2102M

Sept 30, 2021

2103M**

Aug 18, 2022

2105M**

Sep 11, 2022

2106M**

Jan 21, 2022

2107M**

Sept 30, 2022

2109M

Oct 15, 2022

2110M

Oct 29, 2022

2111M

May 31, 2022

2112M

Nov 13, 2022

2113M

Jun 30, 2022

2114M

Dec 17, 2022

7114M

Dec 17, 2022

2115M

Dec 29, 2022

82-0004
70-8098

LeadCare Plus® Blood Lead Test Kit
LeadCare Ultra® Blood Lead Test Kit

2011MU*

Mar 25, 2022

2104MU**

Aug 25, 2022

2108MU**

Mar 31, 2022

* Lots previously included in the recall initiated on May 7, 2021
** Lots previously included in the recall initiated on June 11, 2021
*** Only the following sub-lots of lot 2012M are included in the recall: -08, -09, -10, -11, -12, -13, and -14